Catalog Number 011-MC3322
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
The device is available for evaluation, however, has not yet been received.Additional reporter: (b)(6).
 
Event Description
The event occurred on various dates between (b)(6) 2023 in the intensive care ward.The event involved a 15 cm (6") appx 0.35 ml, smallbore bifuse ext set w/2 microclave¿ clear, 2 clamps, rotating luer where it was reported patients have not received medication due to leakage in the luer lock coupling to the svk/pvk.This occurred during continuous infusions using syringe volume pumps.The defect was visible, and the location of the defect was in the luer lock.The patient/ health care provider were treated with painkillers.After using the product, the cracks were detected.There was patient involvement but no patient harm.
 
Manufacturer Narrative
A series of photos were shared by the customer, where an used opened packaged item #011-mc3322 is observed broken the male luer, no additional damage or anomalies are observed on the photo.One new sample item # 011-mc3322 was returned for evaluation, as received no physical damage or anomalies were visually observed on the sealed product.No mating devices were returned.The sample was tested as per procedure and no leaks, occlusions or anomalies were confirmed after priming.Complaint of leaks cannot be confirmed or replicated.The lot history was reviewed, and no nonconformities were identified that may have contributed to the reported complaint.D9: device returned to manufacturer on 2/19/2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
15 CM (6") APPX 0.35 ML, SMALLBORE BIFUSE EXT SET W/2 MICROCLAVE® CLEAR, 2 CLAMP
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18667346
MDR Text Key334850347
Report Number9617594-2024-00121
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
K964435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number011-MC3322
Device Lot Number13498374
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received02/08/2024
Supplement Dates Manufacturer Received03/06/2024
Supplement Dates FDA Received04/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SYRINGE VOLUME PUMP, UNK MFR
-
-