• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 7300TFX
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombosis/Thrombus (4440)
Event Date 01/16/2024
Event Type  Injury  
Manufacturer Narrative
H10: additional manufacturer narrative: the device wont be available for return because customer sent the device to hospital pathology.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
Edwards received notification that a 7300tfx29mm valve implanted in the mitral position was explanted after approximately two (2) years due to thrombosis leading to regurgitation.Patient was 12 years old.A mechanical valve was implanted in replacement.Per clinical observation, one of the reasons was likely an inadequate anticoagulant therapy because the patient's family were not administering aspirin as directed.
 
Manufacturer Narrative
Added information to section h6 (type of investigation), h6 (investigation findings) and h6 (investigations conclusions).H10: additional manufacturer narrative: the device history record (dhr) was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.The device was not returned to edwards for further investigation and no images or medical records were provided.Thrombosis is a well-recognized complication of prosthetic devices.Device thrombosis is the formation of blood clots on the device/graft.There may be cases of incidental finding by imaging (ct scan) of thicken leaflets (halt) when the patient will benefit from a close follow-up and may be treated with oral anticoagulant.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The most likely cause is patient factors, including potentially inadequate anticoagulant therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS
Type of Device
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 office m2013
irvine, CA 92614
9492506615
MDR Report Key18668193
MDR Text Key334951414
Report Number2015691-2024-00834
Device Sequence Number1
Product Code DYE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number7300TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/16/2024
Initial Date FDA Received02/08/2024
Supplement Dates Manufacturer Received03/08/2024
Supplement Dates FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization; Life Threatening;
Patient Age12 YR
Patient SexFemale
-
-