Model Number UHI-4
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information is provided by the user facility.
 
Event Description
It was reported, the insufflator malfunctioned while switching the device on, the leds light up, and the device switches off and starts beeping.The issue occurred during preparation for use.There was no patient involvement.
 
Manufacturer Narrative
Updated fields: h6 and h10.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation it was confirmed the issue was caused by a faulty printed circuit board.However, the specific cause of the faulty printed circuit board was not established at this time.Olympus will continue to monitor field performance for this device.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the results of the device evaluation.The device was returned and evaluated.During evaluation, it was discovered the display panel was turned off due to a failure of the circuit board (cr board).It is likely the age of the circuit board and the stress from heat and humidity had affected the device¿s performance over time and contributed to the device malfunction.A supplemental report will be submitted upon completion of the investigation or if additional information is provided by the user facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18671247
MDR Text Key336185242
Report Number3002808148-2024-01226
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-