Model Number HL-90
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2023
Event Type  malfunction  
Manufacturer Narrative
D5.Other operator of device: operator of device is unknown.Investigation summary: the device received for investigation had a dented and scratched front cover, broken pole clamp, cracked water tank cover, an outdated pcb and power source, corroded quick connect with the enclosure cracked, and a faded and worn line cord.The tank was filled with water, the temperature check disposable was attached, the line cord was plugged in, and the power switch turned on.The pump was concluded to be very loud, confirming the reported event.The root cause was established to be the pump.As a result, the device was deemed to be beyond economical repair and will be scrapped.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
 
Event Description
It was reported that the unit made a loud sound when powered on.While the device still functioned, it was louder than usual.There was no patient involvement and no patient harm/adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 LOW FLOW - HOTLINE FLUID AND BLOOD WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18671666
MDR Text Key334980865
Report Number3012307300-2024-00639
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2024
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received02/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/1999
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-