Model Number 0800-0543 |
Device Problems
Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/10/2024 |
Event Type
malfunction
|
Event Description
|
A customer reported that during a patient procedure, using a glidescope spectrum smart cable, the video was intermittent due to intermittent connection between the laryngoscope and smart cable.No delay in the procedure, use of a backup device, or harm to the patient was reported.
|
|
Manufacturer Narrative
|
The customer declined the option to have their glidescope spectrum smart cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.However, upon verathon following up with the customer, they reported there being a break in the video cable wiring to the blade connector.It is likely that the reported damage to the smart cable may have caused or contributed to the image issues.The glidescope video laryngoscopes operations and maintenance manual (omm) notes that "before every use, ensure that the instrument is operating correctly and has no sign of damage.Do not use this product if the device appears damaged." verathon followed up with the customer and restated the importance of checking the device before its use in a procedure.Review of complaint history for the reported smart cable serial number " (b)(6)" did not identify any previous complaints.Trending analysis for glidescope spectrum smart cables does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
|
Search Alerts/Recalls
|
|
|