Catalog Number A1059
Device Problem Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: investigation of the returned unit shows that the lock has rotational and lateral movement and a residue buildup is present.New components were added to replace worn internal parts.Also, the skull clamp base has severely worn starburst teeth that will require the clamp base to be replaced.To resolve the issues, the roundhead screw, skull clamp base, roll pins, adjustment screw, label, washers, ball bearings, o-ring and wave springs will be replaced along with general cleaning and maintenance performed.Root cause: the complaint is confirmed via inspection of the unit.The lock was loose and unit required replacement of worn parts and cleaning.The probable root cause is routine use and wear of the unit.No further investigation is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Event Description
A facility reported that the locking mechanism on mayfield modified skull clamp (a1059) was loose.No information has been provided on patient involvement, injury, or surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key18673251
MDR Text Key335323345
Report Number3004608878-2024-00015
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2024
Initial Date Manufacturer Received 01/16/2024
Initial Date FDA Received02/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-