• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT CRM S.R.L. GALI; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT CRM S.R.L. GALI; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRON Back to Search Results
Model Number GALI 4LV SONR CRT-D 2844
Device Problem Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Manufacturer Narrative
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.The atrial lead impedance is stable around 400 ohms on the trends since july 2023 (at least) the high atrial impedance in red on the summary of the device information on the pdf report is linked a coding error in the software code that generates the pdf report.No general malfunction is suspected on the device.This case is retained and utilized for trend purposes.
 
Event Description
Reportedly, on the report, the atrial impedance is displayed>3000 ohm before irm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALI
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRON
Manufacturer (Section D)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT  13040
Manufacturer (Section G)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT   13040
Manufacturer Contact
elodie vincent
via crescentino s.n
saluggia (vc) 13040
IT   13040
146013665
MDR Report Key18675562
MDR Text Key335008515
Report Number1000165971-2024-00119
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGALI 4LV SONR CRT-D 2844
Device Catalogue NumberTDF054C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-