MICROPORT CRM S.R.L. GALI; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRON
|
Back to Search Results |
|
Model Number GALI 4LV SONR CRT-D 2844 |
Device Problem
Data Problem (3196)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/13/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.The atrial lead impedance is stable around 400 ohms on the trends since november 2023 (at least) the high atrial impedance in red on the summary of the device information on the pdf report is linked a coding error in the software code that generates the pdf report.No general malfunction is suspected on the device.This case is retained and utilized for trend purposes.
|
|
Event Description
|
Reportedly, on the report, the atrial impedance is displayed>3000 ohm before irm.
|
|
Search Alerts/Recalls
|
|
|