Items returned for evaluation: delivery system- pusher.Items not returned for evaluation: web implant; introducer; dispenser hoop; microcatheter; controller.The visual analysis of the returned items found the implant to be separated from delivery system and the implant was not returned for evaluation.Inspection of the pusher found the hypotube kinked at the hypotube to connector junction, and the proximal connector bent at the solder joints.Tested the returned device with an in-house controller and gave red lights.The delivery system resistance was measured to be (spec= 66-78, which is out of specification due to the connector damage.The heater coil section was dissected to investigate any obstruction that could possibly prevent the implant from detaching, and the heater coil was found to be stretched which is an indication that the tether could have been caught in between the coil windings and caused the heater coil to stretch when the delivery system was pulled/retracted during the procedure.The heater coil did show signs of controller activation as indicated by the melted tether and pet.The heater coil winding damage likely also contributed to the failed continuity and resistance testing.Investigation conclusion: the investigation of the returned web system found the implant separated from the delivery system, the hypotube kinked, the proximal connector kinked, and the heater coil windings stretched.The implant was not returned for evaluation as it was pushed into the aneurysm as described in the reported event.The device failed continuity and resistance testing, which is consistent with the damage to the proximal connector and the heater coil windings.However, the heater coil pet and tether were found melted, indicating that the device was activated using a detachment controller during the procedure; therefore, the damage to the proximal connector and heater coil windings likely occurred post-activation.The stretched heater coil windings are an indication that the tether could have been caught in between the coil windings and caused the heater coil to stretch when the delivery system was pulled/retracted during the procedure.The physical evaluation of the device could not identify the conditions or circumstances that led to the delivery system damage, but the damage is consistent with the device experiencing forces over specification.
|