Additional information: d4, h3, h4, h6, h10 h4: the lot was manufactured between april 3, 2023 - april 5, 2023.H10: the actual device was received for evaluation.Visual inspection was performed which observed that the bladder has been ruptured.The ruptured bladder was examined for signs of abnormality; no signs of abnormality were observed.The reported condition was verified.The cause of the condition could not be determined; however, a possible cause may be due to inherent variation in the material from manufacturing, as bladders are manufactured with rubber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|