Catalog Number 22-4035 |
Device Problems
Break (1069); Difficult to Remove (1528); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/18/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that during a shoulder arthroscopy, when assembling the clamp of the firstpass suture passer, the rack part of it was damaged (nonetheless, the user was able to assemble and use it).Additionally, the wire was not bringing, it was pulled out with a grasper.The procedure was successfully completed with a delay of less than 30 minutes using the same device.There was no damage to the patient's health.No further complications were reported.
|
|
Manufacturer Narrative
|
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation.However, an image evaluation was performed and found the complaint device, the upper jaw is missing a piece.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since findings cannot lead to a clear cause of the reported event.Factors that could have contributed to the reported event include excessive force, tissue thickness, damage or debris on the device tip between passes.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.Corrected data: h6: medical device problem code.
|
|
Search Alerts/Recalls
|
|