• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW + ADULT NASAL CANNULA Back to Search Results
Model Number OPT944
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product background: the opt944 optiflow + adult nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).Optiflow + interfaces are designed for use with the airvo 2 series of humidification devices and may also be compatible with the fisher and paykel mr850 and 950 humidification system devices.Nhf therapy is intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases nhf therapy should not be used for life support purposes, and appropriate patient monitoring must be used at all times.The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint opt944 optiflow + adult nasal cannula was not returned to f&p for evaluation.Our investigation is based on the information and photograph provided by the healthcare facility, previous investigations of similar complaints, and our knowledge of the product.Results: visual inspection of the provided photograph revealed that the tubing of the opt944 optiflow + adult nasal cannula was detached from the 3-way connector.Conclusion: our investigation was unable to determine the cause of the observed damage to the opt944 optiflow + adult nasal cannula.Based on our knowledge of the product the tubing was likely subjected to excessive force.Manufacturing controls include inspections during production for visual defects to the optiflow + tubing and the swivel, including cracks, tears, moulding defects, contamination, inclusions, discoloration and stretching or deformation.The optiflow + tubing is also inspected for any assembly defects, including confirmation the swivel and 3-way connector are connected with the correct engagement.The subject opt944 optiflow + adult nasal cannula would have met the required specifications.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula, including ensuring the head strap clip and the tubing clip are appropriately attached.The user instructions also state: "do not crush or stretch tube, to prevent loss of therapy." "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." "cannula can become unattached if not used with the head strap clip." "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." "failure to use the set-up described above can compromise performance and affect patient safety.".
 
Event Description
A distributor reported on behalf of a healthcare facility in china that the tubing of an opt944 optiflow + adult nasal cannula was found damaged during use.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + ADULT NASAL CANNULA
Type of Device
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key18682907
MDR Text Key335206384
Report Number9611451-2024-00105
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K162553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/11/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot Number2102191944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/14/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-