• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AUTOSOFT XC; UNO INSET I 60/6 GREY TCAP 10PK INT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AUTOSOFT XC; UNO INSET I 60/6 GREY TCAP 10PK INT Back to Search Results
Lot Number 6002259
Device Problem Material Twisted/Bent (2981)
Patient Problem Diabetic Ketoacidosis (2364)
Event Type  Injury  
Event Description
Unomedical reference number (b)(4).Event occurred in the united states.It was reported that an 11-years-old female child patient faced a kinked cannula due to which she experienced high blood glucose level.Therefore, they tried to treat it with bolus via pump, but on (b)(6) 2023, the patient first went to the emergency room and was subsequently hospitalized due to high blood glucose level.Her highest blood glucose level was 600 mg/dl and ketone level were high which the healthcare professional did not assess as dangerous or life/threatening.Moreover, the issue occurred with one infusion set used for two days.During hospitalization, the patient received fluids of saline, insulin, and unspecified medication (drug name unknown) intravenously as corrective treatment which resolved the issue.On (b)(6) 2023, the patient was released from the hospital with no permanent damage.Unomedical do not see bent/kinking as being related to human factors, but rather as a training issue including correct choices of insertion sites and infusion sets and cannula length.Furthermore, the soft cannula is a flexible material that during use and upon removal can bend slightly.No further information was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSOFT XC
Type of Device
UNO INSET I 60/6 GREY TCAP 10PK INT
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key18683646
MDR Text Key335070279
Report Number3003442380-2024-00144
Device Sequence Number1
Product Code FPA
Combination Product (y/n)Y
PMA/PMN Number
K032854
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Lot Number6002259
Date Manufacturer Received02/06/2024
Patient Sequence Number1
-
-