• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 86 CM; Drug eluting permanent left ventricular (lv) pacemaker electrode

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD DOUBLE BEND, 86 CM; Drug eluting permanent left ventricular (lv) pacemaker electrode Back to Search Results
Model Number 1457Q/86
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2024
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2024-22371.Related manufacturer reference number: 2017865-2024-22372.It was reported that the patient's atrial lead, right ventricular (rv) lead, and left ventricular (lv) leads were all failing to capture.Fluoroscopy was performed and confirmed dislodgement.It was alleged the dislodgement occurred due to twiddler's syndrome.During lead repositioning procedure, the helices of the atrial and rv leads failed to extend.The lv lead was repositioned and then dislodged again upon fixation.The atrial and rv leads were explanted and replaced, and the left ventricular lead was removed and the port was plugged.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD DOUBLE BEND, 86 CM
Type of Device
Drug eluting permanent left ventricular (lv) pacemaker electrode
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18685883
MDR Text Key335133443
Report Number2017865-2024-22373
Device Sequence Number1
Product Code OJX
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1457Q/86
Device Lot NumberA000148031
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age82 YR
Patient SexFemale
Patient Weight73 KG
-
-