• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0706-XTW
Device Problem Incomplete Coaptation (2507)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2024
Event Type  Injury  
Event Description
It was reported this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4+, flail and pre-existing ruptured papillary muscle.One xtw clip was successfully implanted.To further reduce mr, an additional xtw clip was inserted and deployed on the mitral valve.However, after deployment, the second clip detached from the posterior leaflet and remained attached to the anterior leaflet (single leaflet device attachment/slda).To stabilize the slda, an additional clip was implanted, reducing mr to a grade of 3.There was no clinically significant delay in the procedure.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history did not indicate a lot-specific product issue.All available information was investigated, the reported slda associated with the clip detaching from the posterior leaflet appears to be due to patient conditions (anterior prolapse and flail extending medially and a lot of motion on the anterior leaflet due to the billowing prolapsing segment and ruptured papillary muscle).Unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDR)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18691506
MDR Text Key335187357
Report Number2135147-2024-00620
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberCDS0706-XTW
Device Lot Number30913R1017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
IMPLANTED MITRACLIP; STEERABLE GUIDE CATHETER
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient SexMale
-
-