• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST; MARYLAND BIPOLAR FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST; MARYLAND BIPOLAR FORCEPS Back to Search Results
Model Number 471172-17
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Event Description
It was reported that during a da vinci-assisted surgical procedure, the maryland bipolar forceps (mbf) instrument pulley cover had a thermal damage.The customer used a backup mbf instrument to continue with the planned procedure.No fragment fell inside the patient.No known impact or patient consequence was reported.Intuitive surgical, inc.(isi) followed up with the site and obtained the following additional information: the mbf instrument was inspected prior to use with no issue.The customer observed charred tissue on the instrument.There was no report of the tissue being burned unexpectedly.The instrument did not collide with any energy instrument during the procedure.There was no arcing observed during the procedure.There was no patient injury.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) has not received the maryland bipolar forceps instrument for evaluation.Therefore, the root cause of the customer reported failure mode has not been determined.A follow-up mdr will be submitted if the product is returned and evaluated and/ or if additional information is received.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the maryland bipolar forceps instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the reported complaint.The instrument was found to have charring and localized melting at the grip base between the grips.The instrument passed the electrical continuity test.The instrument was placed on an in-house system, and no arcing was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
MARYLAND BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18691798
MDR Text Key335184129
Report Number2955842-2024-11249
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471172-17
Device Catalogue Number471172
Device Lot NumberK12230608 0058
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/12/2024
Supplement Dates Manufacturer Received02/22/2024
Supplement Dates FDA Received03/11/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-