Catalog Number 1000-00-101
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2024
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event Description
It was reported from israel that during pre-surgery, it was discovered that the impactor device was working without being activated.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in (b)(6) 2024.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device had corrosion/rusting/pitting and would not run.It was further determined that the device failed pretest for visual assessment, impactor operation assessment, intermittent test assessment and final assessment.The assignable root cause was determined to be traced to component failure from wear and maintenance, which is improper maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINCISE¿ AUTOMATED SURGICAL IMPACTOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
TECHTRONIC INDUSTRIES MEDICAL
1428 pearman dairy rd
anderson SC
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key18694713
MDR Text Key335246331
Report Number1045834-2024-00162
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Followup
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1000-00-101
Device Lot NumberJJAA11509
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received02/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2023
Type of Device Usage Unknown
Patient Sequence Number1
-
-