• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ILIVIA 7 VR-T DF4 PROMRI; ICD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ILIVIA 7 VR-T DF4 PROMRI; ICD Back to Search Results
Model Number 404626
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  malfunction  
Event Description
It was reported that the device shows the eri status.No adverse patient events were reported.Device currently remains implanted.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
The device was received and analyzed.Explant date is currently unknown.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.Next, the device was interrogated.The interrogation could be properly performed and revealed the eri battery status.Next, the icd was subjected to an analysis of the electrical parameters.The current consumption of the icd was checked and found to be normal and as expected.Analysis of the battery condition, however, showed an inconsistency of charge taken from the battery and the battery voltage.A premature battery depletion could be confirmed during analysis.Please note, this icd is affected by the field safety corrective action, bio-lqc, initiated in march 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILIVIA 7 VR-T DF4 PROMRI
Type of Device
ICD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18695691
MDR Text Key335255514
Report Number1028232-2024-00805
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
PMA/PMN Number
P050023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Followup
Report Date 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Model Number404626
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/13/2024
Supplement Dates Manufacturer Received05/03/2024
Supplement Dates FDA Received05/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-