Model Number 471172-17 |
Device Problem
Thermal Decomposition of Device (1071)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A return material authorization (rma) was issued to evaluate the maryland bipolar forceps instrument.A follow-up mdr will be submitted if the instrument is returned (post failure analysis evaluation) or if additional information is received.
|
|
Event Description
|
It was reported that during a da vinci-assisted radical prostatectomy without lymphadenectomy surgical procedure, the maryland bipolar forceps instrument had a melted wrist.The user completed the procedure with no further issue reported.No fragment fell inside the patient.No known impact or patient consequence was reported.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the instrument was inspected prior to use with no damage observed.The thermal damage was first identified during the use of the instrument.The surgeon was unsure what caused the thermal damage.There was no instrument collision and no arcing observed.There was no injury to the patient.
|
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) received the maryland bipolar forceps instrument to perform failure analysis.The maryland bipolar forceps instrument was analyzed and found to have charring and localized melting on the bipolar yaw pulley.The thermal damage was found at the base of the yaw pulley near the grip.The electrode was exposed as a result.The instrument passed an electrical continuity test.The instrument was placed and driven on an in-house system and delivered energy with no signs of arcing.
|
|
Event Description
|
Refer to h10/h11 for follow-up information.
|
|