Model Number 471172-17
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2023
Event Type  malfunction  
Manufacturer Narrative
A return material authorization (rma) was issued to evaluate the maryland bipolar forceps instrument.A follow-up mdr will be submitted if the instrument is returned (post failure analysis evaluation) or if additional information is received.
 
Event Description
It was reported that during a da vinci-assisted radical prostatectomy without lymphadenectomy surgical procedure, the maryland bipolar forceps instrument had a melted wrist.The user completed the procedure with no further issue reported.No fragment fell inside the patient.No known impact or patient consequence was reported.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: the instrument was inspected prior to use with no damage observed.The thermal damage was first identified during the use of the instrument.The surgeon was unsure what caused the thermal damage.There was no instrument collision and no arcing observed.There was no injury to the patient.
 
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the maryland bipolar forceps instrument to perform failure analysis.The maryland bipolar forceps instrument was analyzed and found to have charring and localized melting on the bipolar yaw pulley.The thermal damage was found at the base of the yaw pulley near the grip.The electrode was exposed as a result.The instrument passed an electrical continuity test.The instrument was placed and driven on an in-house system and delivered energy with no signs of arcing.
 
Event Description
Refer to h10/h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
MARYLAND BIPOLAR FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18699304
MDR Text Key335278396
Report Number2955842-2024-11140
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K214095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number471172-17
Device Catalogue Number471172
Device Lot NumberK10230615 0212
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES