The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the control body fluid damage, the light guide cable fluid damage, the lg connector fluid damage, the lcb (light carrying bundle) broken, the remote control buttons cracked, the control body corroded, the mechanical base plate corroded, the lg connector corroded, the shield cover corroded, the remote control buttons leak, the remote control buttons cut, the distal body worn out, and the angle wire grinding; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|