A non-health care professional reported that during an intraocular lens (iol) implant procedure, reported with a description of lens remains trapped in the incision, viscoelastic was placed and attempts are made to insert it but it does not enter.It was extracted outwards, marks are seen on the lens and it is replaced with another lens of the same power.Lens was explanted at initial procedure.Additional information was received.There was no patient harm.
|
Correction: on initial mdr the product code of was an error.It should have been a050302 on the original mdr.The product was returned for analysis and damage was observed to the iol.The device was returned in the blister tray.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced out of the device.The lens is returned adhered to the device with tape.A significant amount of solution/residue is dried on both surfaces of the optic and haptics.Both haptics are intact.The optic is cracked/fractured and scratched/marked-rejectable.The product investigation could not identify the root cause for the reported complaint "lens stuck in the incision, marks are seen on the lens" the lens was returned outside of the device.Lens damage may occur: ¿ due to the use of a non-qualified viscoelastic.Lens delivery performance may be negatively affected when using other non-qualified viscoelastic leading to underfill, overfill, misfolding , delivery issues and /or damage.¿ if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly and cause delivery issues and /or damage.¿ if the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement and cause delivery issues and / or product damage.Any of the above listed causes alone, or in combination, may create the reported event.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.Based on these investigation findings, we are unable to verify if the iol/device contributed to the event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|