Model Number N/A |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/21/2024 |
Event Type
malfunction
|
Event Description
|
It was reported during a bipolar hip replacement, a hair was noticed inside the sterile bag that the head comes in.The issue was found before taking the head out of the bag.The nurse passed it off and got another head of the same size/offset.No consequences or impact to the patient.
|
|
Manufacturer Narrative
|
(b)(4).G2: foreign: canada.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional and/or corrected information.Visual evaluation of the returned product found the packaging was previously opened.A hair-like fiber was found in the sterile packaging.As the product was returned opened, the source of the debris cannot be confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history found no additional related issues for the reported part and part lot combinations.Medical records were not provided.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was returned opened.This complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|