• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Catalog Number FEM10120
Device Problems Positioning Failure (1158); Fracture (1260); Misfire (2532)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/22/2024
Event Type  malfunction  
Manufacturer Narrative
H10: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation; the stent graft was found partially deployed and the outer sheath was fractured which leads to confirmed results.It was reported that a 9fr introducer / 0.035" guidewire were used for access, there was no tortuosity but the vessel had blood clots, pre-dilatation was performed and the device was flushed before use.Based on the available information and evaluation of the returned sample, the investigation is closed with confirmed results for partial deployment and sheath fracture.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.The instructions for use state: 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.' regarding preparation of the device the instructions for use state that "prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment".Regarding the anatomy of the placement site the instructions for use states: "prior to stent graft deployment (.), ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy".Regarding accessories the instructions for use states: "prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location.The use of an appropriately sized introducer sheath is recommended".The packaging pictograms indicate the use of an 9f introducer.H10: d4 (expiration date: 08/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event Description
It was reported that during a stent graft placement procedure in the brachial cephalad vein via the brachial vein access, the device allegedly did not deploy.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Type of Device
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18704710
MDR Text Key335715591
Report Number2020394-2024-00201
Device Sequence Number1
Product Code PFV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberFEM10120
Device Lot NumberANHV0434
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/08/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
Patient Weight86 KG
Patient EthnicityHispanic
-
-