Model Number 799-AEB1-00-01
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Complainant alleged that during a routine shift check by a clinician, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Manufacturer Narrative
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing, which included bench handling, power cycling, and functional stress testing without duplicating the customer's report.The main battery and coin battery were replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI-VENT, MODEL 731
Type of Device
VENTILATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
scott august
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key18704755
MDR Text Key335932737
Report Number1220908-2024-00506
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00847946024192
UDI-Public00847946024192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number799-AEB1-00-01
Device Catalogue Number799-AEB1-00-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/25/2024
Initial Date FDA Received02/14/2024
Supplement Dates Manufacturer Received01/25/2024
Supplement Dates FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-