Device evaluated by manufacturer: the returned product consisted of the synergy xd mr ous 5.00 x 12mm device, tracked over a guidewire and inserted through a guide catheter.A visual, tactile microscopic and device to device interaction examination of the returned device was performed.Device analysis confirmed that the device was stretched down onto the guidewire which prevented the movement of the device over the wire.Just the distal section of the balloon and tip was exiting from the tip of the guide catheter.No damage was observed to the balloon however, the balloon was not tightly folded.The tip did not exhibit any signs of damage however, the inner/wire lumen was stretched.The device could not be removed or advanced through the guide.Attempts to pull a vacuum to try and tighten the balloon folds failed, consistent with stretching in the distal extrusion.The device was trapped in the guide catheter.
|