The device was returned for analysis.The reported hub break and hub leak were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It was reported that the device was not prepped (air aspirated) prior to use.It should be noted that the omnilink elite instructions for use, states: prepare an inflation device / syringe with diluted contrast medium.Attach an inflation device / syringe to the stopcock; attach to the inflation port.With the tip down, orient the delivery system vertically.Open the stopcock to delivery system; pull negative for 30 seconds; release to neutral for contrast fill.Close the stopcock to the delivery system; purge the inflation device / syringe of all air.Repeat steps until all air is expelled.In this case, it is unlikely the ifu deviation related to incorrect prep contributed to the reported event.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|