Model Number RSINT40022X
Device Problems Material Deformation (2976); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2024
Event Type  malfunction  
Event Description
An attempt was made to use one resolute integrity coronary drug eluting stent to treat a moderately tortuous, moderately calcified lesion with 80% stenosis in the mid left main (lm) coronary artery.The device was inspected with no issues noted.The device did not pass through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.It was reported that the device failed to cross the lesion.It is believed that the event was due to use of the device in difficult lesion morphology/anatomy, i.E.The event was procedural related and not device related.The patient is alive with no injury.
 
Manufacturer Narrative
Product analysis: the device was received for analysis.The stent was positioned on the balloon between the marker bands as per position specification, but due to mid and distal stent deformation, the stent did not meet visual acceptance specification.Deformation was evident to the distal stent wraps with struts raised and bunched.A kink was evident to the mid stent wraps.No deformation was evident to the distal tip.The inner lumen patency was verified with a 0.015-inch mandrel.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE INTEGRITY RX
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18708996
MDR Text Key335443879
Report Number9612164-2024-00787
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
P110013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRSINT40022X
Device Catalogue NumberRSINT40022X
Device Lot Number0011514448
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2024
Initial Date Manufacturer Received 02/09/2024
Initial Date FDA Received02/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial