Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) batteries is not charging.There was no patient involvement reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate the issue.Upon arrival onsite, the fse found the unit plugged in and the batteries with no charge and the console not fully engaged.Once the console was reinserted into the cart, the iabp powered on and the batteries began charging.The unit passed all calibration, functional and safety tests per factory specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18710251
MDR Text Key335413253
Report Number2249723-2024-00652
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-