• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXFORD UNI CMNTLS TIB SZ C LM; OXFORD CEMENTLESS TIBIAL TRAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXFORD UNI CMNTLS TIB SZ C LM; OXFORD CEMENTLESS TIBIAL TRAYS Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Insufficient Information (4580)
Event Date 02/06/2024
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10 - medical devices.Oxford ph3 cementless fem sz m; item# 154926; lot# 7074536.Unknown bearing; item# unknown; lot# unknown.G2 - foreign: australia.Multiple mdr reports were filed for this event, please see associated reports: 3002806535 - 2024 - 00058.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that a patient had a cementless initial left unicompartmental knee replacement performed.Subsequently, the patient was revised due to unspecified symptoms, vague hot spots identified on ct, and potential infection.During the revision, the tibial component was replaced with a cemented tibia, and the bearing was also exchanged.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional and corrected information.The product has been discarded and was not available for investigation.Visual examination of the provided pictures identified only little evidence of bony ongrowth on the tibial tray which was implanted for approximately eighteen months before revision surgery.The bearing does not appear to be damaged or worn, however based on the quality of the provided photograph a conclusive assessment cannot be determined.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Medical records were provided and reviewed by a health care professional.The review identified that the information provided is insufficient and inconclusive regarding symptoms, allegations, and findings upon revision.Infection is suspected but not confirmed and cannot be correlated with vague hot spots found on ct.Recommend serious injury at this time due to insufficient information as symptoms necessitated revision surgery in which the initial device cannot be ruled out.The hot spots found on ct scans could represent infection or loosening.The visual examination identified only little evidence of bony ongrowth on the tibial tray which could possibly point to loosening.However, with the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD UNI CMNTLS TIB SZ C LM
Type of Device
OXFORD CEMENTLESS TIBIAL TRAYS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18710549
MDR Text Key335464321
Report Number3002806535-2024-00057
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P010014/S078
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number166574
Device Lot Number7126610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/06/2024
Initial Date FDA Received02/15/2024
Supplement Dates Manufacturer Received03/18/2024
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
Patient SexFemale
-
-