• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AUTOCAT 2; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AUTOCAT 2; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0400
Device Problem Failure to Run on Battery (1466)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).N/a.Other remarks: n/a.Corrected data: n/a.
 
Event Description
It was reported that during transportation of a patient on battery power, battery power failed and the pump shut down.Additional in formation received states that the patient did not suffer any injuries.At the time of this report, the customer has not returned our requests for additional information.If further information is received, the complaint file will be updated.
 
Event Description
It was reported that during transportation of a patient on battery power, battery power failed and the pump shut down.Additional in formation received states that the patient did not suffer any injuries.At the time of this report, the customer has not returned our requests for additional information.If further information is received, the complaint file will be updated.
 
Manufacturer Narrative
(b)(4).The reported complaint of short battery runtime was not able to be confirmed as no part or recorder strip was returned for investigation.Iabp users and servicers must follow the operating instructions manual for recommendations on usage, charging, maintenance and storage of the batteries.If battery maintenance is not performed per the iabp operating instructions manual, the battery may not provide the expected minimum run time of operating power.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint was undetermined.No further action was required at this time.Teleflex will continue to monitor and trend on complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT 2
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville 27560
MDR Report Key18711922
MDR Text Key335536855
Report Number3010532612-2024-00119
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIAP-0400
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/01/2024
Initial Date FDA Received02/15/2024
Supplement Dates Manufacturer Received03/21/2024
Supplement Dates FDA Received03/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-