SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Catalog Number 72200616 |
Device Problems
Mechanical Problem (1384); Overheating of Device (1437); Key or Button Unresponsive/not Working (4063)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the mdu buttons were not working, there was an error coming up saying blades stalled handpiece started to get warm.No case reported; therefore, there was no patient involved.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
Internal complaint reference (b)(4).The reported device was received for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.Unresponsive buttons, mechanical jam and overheating have been thoroughly investigated in similar complaints and are anticipated within the risk files and the instructions for use, therefore a product evaluation was not performed.The root cause was associated with a mechanical component failure based on analysis of similar complaint events.Factors that can contribute to unresponsive buttons include failure of the reed switch or damage on the hall board which may contribute to a short circuit.Factors that can contribute to mechanical jam and overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|