Catalog Number 1120350-28
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Angina (1710); Stenosis (2263)
Event Date 12/30/2022
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.A conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined; however, the subsequent treatment appears to be related to the operational context of the procedure.The reported patient effects of angina and stenosis are listed in the xience alpine everolimus eluting coronary stent system (eifu), electronic instructions for use as a known patient effects of coronary stenting procedures.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that on (b)(6) 2022, the patient had percutaneous transluminal coronary angioplasty (ptca) in the left main (lm) artery to the left anterior descending (lad) and the left circumflex artery.A 3.5x28mm xience alpine stent was successfully implanted.On (b)(6) 2024 the patient presented with chest pain and angiography was performed.In-stent restenosis was noted in the lm to lad artery in the mid segment of the stent.On (b)(6) 2024, three cutting balloons and three drug-eluting balloons were used to treat the restenosis.There was no adverse patient sequela.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18713356
MDR Text Key335522789
Report Number2024168-2024-01882
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number1120350-28
Device Lot Number2070141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2024
Initial Date FDA Received02/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-