Visual analysis was performed on the returned product.The reported premature activation was confirmed as the filter was returned outside of the delivery catheter.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history revealed no indication to suggest a lot specific product quality issue exists.The investigation was unable to determine a cause for the reported premature activation.It may be possible that during advancement, the noted damage to the delivery catheter pod occurred resulting in the filter deploying prematurely; however, this could not be confirmed.Additional information received from the account stated the break in the filter support structure may have occurred at the end of the procedure, as the customer was packaging the device; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|