SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Catalog Number 72200616 |
Device Problems
Corroded (1131); Overheating of Device (1437); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/26/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during a shoulder arthroscopy, the mdu was running hot.The procedure was completed with a smith and nephew back up device and it is unknown if there was a delay.No further complications were reported.
|
|
Manufacturer Narrative
|
H10: internal complaint reference: (b)(4).
|
|
Manufacturer Narrative
|
H2: additional information: h6: medical device problem code.H3, h6: the reported device was received for evaluation.A visual inspection revealed a discolored cable.Discolored cable is a cosmetic issue that does not affect device functionality and it is not related to the alleged failure.A functional evaluation was performed on the returned device and found a noisy motor.Further evaluation revealed a corroded gearbox.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with a component failure.Factors that could have contributed to the failure include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|