BAXTER HEALTHCARE CORPORATION CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 2C8634 |
Device Problems
Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/18/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported a clear link system extension set leaked.The tube appeared to be completely severed at the junction.Blood was observed on the patient¿s gown and the floor.The volume of blood loss was not known.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.Visual inspection was performed which identified a cut in the tube.The reported condition was verified.The cause of the condition could not be determined; however, a potential cause is a cut due to a sharp surface during the use of the product.Should additional relevant information become available, a supplemental report will be submitted.
|
|