• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM Back to Search Results
Catalog Number 22438-19
Device Problems Retraction Problem (1536); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2024
Event Type  Injury  
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event Description
It was reported that the procedure was to treat a lesion in the right superficial femoral artery (sfa) with heavy calcification.The emboshield nav 6 embolic protection system (eps) was used, however, the filter got stuck upon partial retrieval.The eps was removed from the sheath, however, the filter separated and a huge chunk of the calcium was in the filter.Surgical cutdown was performed to remove the filter from the anatomy.There was no adverse patient sequelae and there was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Visual analysis was performed on the returned product.The reported retraction problem could not be confirmed due to the condition of the returned unit.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints reported from this lot.The investigation determined that the reported difficulties resulting in surgical intervention appear to be related to circumstances of the procedure.Based on the reported information, it is likely that the difficulty retrieving the filter was due to a large embolic load causing difficulty pulling the filter into the retrieval catheter resulting in damage to the filtration element requiring surgical intervention to remove the filter from the anatomy.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the emboshield nav6 embolic protection device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOSHIELD NAV6 EMBOLIC PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18723249
MDR Text Key335601377
Report Number2024168-2024-01911
Device Sequence Number1
Product Code NTE
UDI-Device Identifier08717648137754
UDI-Public(01)08717648137754(17)261031(10)3111462
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141678
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number22438-19
Device Lot Number3111462
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient SexMale
Patient Weight70 KG
-
-