• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Device Dislodged or Dislocated (2923); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2024
Event Type  malfunction  
Event Description
It was reported that the left ventricular (lv) lead became dislodged during the implant procedure.The lead was explanted and replaced to resolve the event and the patient was in stable condition.Following explant, the lv lead was observed to be twisted.
 
Manufacturer Narrative
As received, a complete lead was returned in one piece.The reported event of twisted lead was not confirmed.A visual examination of the lead revealed no anomalies.Electrical testing was performed and revealed no indication of conductor fractures or internal shorts.The s-curve was measured to be within required performance specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18725556
MDR Text Key335613611
Report Number2017865-2024-32883
Device Sequence Number1
Product Code NIK
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberS000091288
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/29/2024
Initial Date FDA Received02/16/2024
Supplement Dates Manufacturer Received03/04/2024
Supplement Dates FDA Received03/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-