Device Problems
Device Dislodged or Dislocated (2923); Material Twisted/Bent (2981)
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2024
Event Type
malfunction
Event Description
It was reported that the left ventricular (lv) lead became dislodged during the implant procedure.The lead was explanted and replaced to resolve the event and the patient was in stable condition.Following explant, the lv lead was observed to be twisted.
Manufacturer Narrative
As received, a complete lead was returned in one piece.The reported event of twisted lead was not confirmed.A visual examination of the lead revealed no anomalies.Electrical testing was performed and revealed no indication of conductor fractures or internal shorts.The s-curve was measured to be within required performance specification.