• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED PTY LTD ASTRAL 150 - AMER; VENTILATOR, CONTINUOUS (FACILITY/HOME)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED PTY LTD ASTRAL 150 - AMER; VENTILATOR, CONTINUOUS (FACILITY/HOME) Back to Search Results
Model Number 27003
Device Problem Inappropriate Tactile Prompt/Feedback (4023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2024
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed.Evaluation confirmed the reported complaint.The top case was removed and reinstalled to address the issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Type of Device
VENTILATOR, CONTINUOUS (FACILITY/HOME)
Manufacturer (Section D)
RESMED PTY LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key18726363
MDR Text Key335736726
Report Number3007573469-2024-00035
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/16/2024
Distributor Facility Aware Date01/17/2024
Device Age93 MO
Date Report to Manufacturer02/16/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/13/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-