U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Device Problems Fracture (1260); Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2024
Event Type  Injury  
Event Description
It was reported that the stent partially deployed and fractured.A 7x150 eluvia drug-eluting vascular stent system was selected for use.A contralateral approach was used to access the mid superficial femoral artery (sfa) that was severely diseased but not very tortuous.A 6x150 stent was previously placed in the distal sfa.The delivery system was advanced over a 0.014 guidewire.After deploying approximately 60mm of the 7x150 eluvia stent using the thumbwheel, the delivery system failed.The stent was partially deployed in the mid-sfa.The physician then pulled back the blue handle to try and unsheath the remaining 90mm to no avail.The physician subsequentially pulled the entire device back, hoping the stent would deploy.It was at that point the stent fractured into 2 pieces.The stent was unable to be retrieved since a wire was no longer able to pass distally.It was at that point they decided to convert to a bypass.The bypass was completed successfully, and the patient is recovering well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18726836
MDR Text Key335694340
Report Number2124215-2024-09382
Device Sequence Number1
Product Code NIU
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P180011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24657
Device Catalogue Number24657
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-