Model Number VH-4000
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2024
Event Type  malfunction  
Event Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro 2 wire in the jaw separated from the jaw insulation.All materials from the jaw was accounted for, and none left in the tunnel.A second device was used to complete the case.There was no harm to the patient, or delays reported.
 
Manufacturer Narrative
(b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
N/a.
 
Manufacturer Narrative
Trackwise#: (b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h10.The device was returned to the factory for evaluation on 01/30/2024.An investigation was conducted on 01/31/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood and charred material were observed.A microscopic inspection was conducted.The clear silicone insulation of both the hot and cold jaws was observed to be intact with no visual defects.The heater wire was observed to be flexed away from the base of the hot jaw and detached from the tip of the jaw.Based on the returned condition of the device and evaluation results, the reported failure "material twisted/bent; wire" was confirmed.The lot # 3000350756 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW HEMOPRO 2
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key18726954
MDR Text Key336641920
Report Number2242352-2024-00105
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
PMA/PMN Number
K101274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVH-4000
Device Catalogue NumberVH-4000
Device Lot Number3000350756
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2024
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN.
Patient Age55 YR
Patient SexMale