The hospital reported that during an endoscopic vein harvesting procedure, vasoview hemopro 2 wire in the jaw separated from the jaw insulation.All materials from the jaw was accounted for, and none left in the tunnel.A second device was used to complete the case.There was no harm to the patient, or delays reported.
|
Trackwise#: (b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h10.The device was returned to the factory for evaluation on 01/30/2024.An investigation was conducted on 01/31/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood and charred material were observed.A microscopic inspection was conducted.The clear silicone insulation of both the hot and cold jaws was observed to be intact with no visual defects.The heater wire was observed to be flexed away from the base of the hot jaw and detached from the tip of the jaw.Based on the returned condition of the device and evaluation results, the reported failure "material twisted/bent; wire" was confirmed.The lot # 3000350756 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|