Trackwise#: (b)(4).The device was returned to the factory for evaluation on 02/06/2024.An investiagation was conducted on 02/07/2024.A visual inspection was conducted.Signs of clinical use and evidence of blood and charred material was observed.There were no visual defects observed on the intact heater wire or the clear silicone insulation on both the cold and hot jaws.An electrical evaluation was conducted.An electrical evaluation was conducted.A pre-cautery test was performed per the direction for use (dfu) with a reference cable, adapter, and reference power supply vh-3010 at level 3.0.The device passed the pre-cautery test; it produced visible steam and heat during ten (10) 3-second activations and shut off when the toggle was released.No excessive smoke and/or steam were observed during the testing.To evaluate the safety shut down system, a polyfuse activation test was performed 5 times over 10 minutes.The device shut off after the period of sustained activation and reactivated after 10-second cooling period with no incident each time.An activation and transection capability test was performed over four (4) repetitions using "max life test method stm2048073 rev aa.The device successfully transected tissue four (4) times.A temperature and resistance test was conducted to evaluate the device function per hemopro 2 final test 90523436 rev w.The resistance value was measured at.69 ohms which is within specification.The device passed the temperature measurements test.The displayed temperature increased and turned green within the 2 second specified timeframe.The displayed temperature decreased once the toggle swivel was released.Based on the results of the evaluation, the reported failure "failure to cut" was not confirmed.The lot # 3000353124 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|