Model Number CDVRA600Q
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2024
Event Type  malfunction  
Event Description
During a procedure, it was difficult to tighten the lead pin into the header.The device was explanted and replaced.The patient was stable.
 
Manufacturer Narrative
The reported event of a set screw and connection issue was confirmed.The ventricular set screw hex cavity was found to be stripped.The stripped set screw hex cavity prevented the field from properly connecting their leads which caused the reported field events.The cause of the stripped set screw hex cavity was procedure related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUTRINO NXT VR
Type of Device
NO MATCH
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18728212
MDR Text Key335691115
Report Number2017865-2024-22512
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCDVRA600Q
Device Lot NumberP000187095
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
-
-