A field service engineer (fse) conducted a site visit and was able to confirm the problem by comparing the results to another instrument.The fse checked the substrate and wash and looked for signs of leaks and no issues were noted.The fse checked the detector lens and found that the lens was dirty.The fse cleaned the detector lens and ran controls.The resolution was verified by calibrating and running quality control (qc).No further action is required by field service.The aia-900 analyzer is functioning as expected.A complaint history review and service history review from the install date through the aware date of the event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the searched period.The st prog iii analyte application manual states the following: limitations of the procedure for diagnostic purposes, the results obtained from this assay should be used in conjunction with other data (e.G.Symptoms, results of other tests, clinical impressions, therapy, etc.).Using st aia-pack progiii, the highest measurable concentration of progesterone in specimens without dilution is 40 ng/ml, and the lowest measurable concentration in specimens is 0.1 ng/ml (assay sensitivity).Although the approximate value of the highest calibrator is 45 ng/ml, the exact concentration may be slightly different.The assay specification, assay range high, should be defined as the upper limit of the assay range, 40 ng/ml.Although hemolysis has an insignificant effect on the assay, hemolyzed samples may indicate mistreatment of a specimen prior to assay and results should be interpreted with caution.Lipemia has an insignificant effect on the assay except in the case of gross lipemia where spatial interference may occur.Samples from patients who had an injection of fluorescein, which is used in fluorescein fundus angiography, may cause falsely elevated results certain medications may interfere with assay performance.Specimens from patients taking medicines and/or medical treatment may show erroneous results.All results should be interpreted with respect to the clinical picture of the patient.Heterophile antibodies are known to interfere in assays of this type.St aia-pack progiii has been designed to minimize the effects of heterophile antibodies, but interference from high titers cannot be ruled out.For a more complete understanding of the limitations of this procedure, please refer to the specimen collection and handling, warnings and precautions, storage and stability, and procedural notes sections in this insert sheet.The most probable cause of the reported event was a dust or dirt problem, cause traced to environment.
|