TOSOH HI-TEC, INC. AIA-900; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-900
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/22/2024
Event Type  malfunction  
Event Description
A customer reported discrepant results for progesterone (prog iii) on the aia-900 analyzer.The patient physician questioned the results and sent patient samples to a reference lab that is using a roche e2600 for retesting.The customer stated all quality control (qc) results were within acceptable range and no calibration issues.The customer replaced the wash and diluent and reran samples but error persisted.A technical support specialist (tss) sent prog iii sample diluting solution (sds) to the customer and the customer retested samples again using the sds in triplicate and error persisted.The analyzer is down.There was no indication of any patient intervention or adverse health consequences due to the reported event.
 
Manufacturer Narrative
A field service engineer (fse) conducted a site visit and was able to confirm the problem by comparing the results to another instrument.The fse checked the substrate and wash and looked for signs of leaks and no issues were noted.The fse checked the detector lens and found that the lens was dirty.The fse cleaned the detector lens and ran controls.The resolution was verified by calibrating and running quality control (qc).No further action is required by field service.The aia-900 analyzer is functioning as expected.A complaint history review and service history review from the install date through the aware date of the event for similar complaints was performed for serial number (b)(6).There were no similar complaints identified during the searched period.The st prog iii analyte application manual states the following: limitations of the procedure for diagnostic purposes, the results obtained from this assay should be used in conjunction with other data (e.G.Symptoms, results of other tests, clinical impressions, therapy, etc.).Using st aia-pack progiii, the highest measurable concentration of progesterone in specimens without dilution is 40 ng/ml, and the lowest measurable concentration in specimens is 0.1 ng/ml (assay sensitivity).Although the approximate value of the highest calibrator is 45 ng/ml, the exact concentration may be slightly different.The assay specification, assay range high, should be defined as the upper limit of the assay range, 40 ng/ml.Although hemolysis has an insignificant effect on the assay, hemolyzed samples may indicate mistreatment of a specimen prior to assay and results should be interpreted with caution.Lipemia has an insignificant effect on the assay except in the case of gross lipemia where spatial interference may occur.Samples from patients who had an injection of fluorescein, which is used in fluorescein fundus angiography, may cause falsely elevated results certain medications may interfere with assay performance.Specimens from patients taking medicines and/or medical treatment may show erroneous results.All results should be interpreted with respect to the clinical picture of the patient.Heterophile antibodies are known to interfere in assays of this type.St aia-pack progiii has been designed to minimize the effects of heterophile antibodies, but interference from high titers cannot be ruled out.For a more complete understanding of the limitations of this procedure, please refer to the specimen collection and handling, warnings and precautions, storage and stability, and procedural notes sections in this insert sheet.The most probable cause of the reported event was a dust or dirt problem, cause traced to environment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key18728604
MDR Text Key336623669
Report Number3004529019-2024-00513
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/22/2024
Initial Date FDA Received02/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-