MEDTRONIC MED REL MEDTRONIC PUERTO RICO ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37612 |
Device Problems
Failure to Deliver Energy (1211); Battery Problem (2885); Charging Problem (2892); Connection Problem (2900); Material Integrity Problem (2978)
|
Patient Problems
Ambulation Difficulties (2544); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Section d10 references: product id 37761, serial# (b)(6).Product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 37761, serial/lot #: (b)(6).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that patient is not able to charge their recharger.Additional information received from the consumer reported the recharger wasn¿t charging due to the connector pin being pulled off.The implant was ¿completely flat/off.¿ it was mentioned the consumer had dementia and this morning they couldn¿t walk.
|
|
Manufacturer Narrative
|
H3.Analysis of the desktop charger (s/n (b)(6)) found the cord was frayed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information received from the consumer reported the cause of the implant depleting was due to the charging unit damage and thus the charging unit was depleted.A new unit was ordered and after recharging with a loaner, the issue of the inability to walk was resolved.
|
|