An event of valve being calcified with high aortic stenosis was reported.It was also reported that the patient was in cardiogenic shock.A more comprehensive assessment, including histopathological examination of the valve tissue could not be performed as the device was not returned for analysis.However, based on the information received, the reported stenosis is consistent with structural valve deterioration (svd).A variety of factors may contribute to svd, including patient, biological, and implant related factors: no implant related factors could be confirmed from the information received from the field as information related to implant procedure was not provided.Biological factors which can result in tissue degeneration such as calcification (from patient conditions predisposing to elevated calcium like renal disease etc.) or thinning of the prosthetic leaflet tissue (predisposing to leaflet tears) also could not be confirmed as the valve was not returned for histopathological examination.The reported calcifications on valve could have contributed to the reported stenosis, however it could be confirmed as the valve was not received for examination.
|
It was reported that on an unknown date in 2015, a 25mm trifecta valve was implanted.On (b)(6) 2024, a 23mm navitor valve with a small flexnav delivery system was chosen for off-label valve in valve implantation into the trifecta valve in an emergency procedure.The trifecta valve was calcified with high aortic stenosis.The patient was in cardiogenic shock.The navitor was implanted successfully within the trifecta with a good result.No additional information was provided.There was no patient consequences.
|