D10 concomitant product: lxmj44s, maryland xp inline sealer/divider 44cm (lot#unknown); vlft10gen, vlft10gen ft series energy platformx1 (serial#unknown) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the jaws were difficult to open, the knife blade did not advance or difficult to advance, the device was difficult/impossible to close and the knife blade did not retract.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
According to the reporter, during single anastomosis duodenal-ileal bypass with sleeve gastrectomy, about 0-2 seals, of less than 0.2 mm thickness in omentum, the device would not cut properly leaving jagged edges.Patient experienced more bleeding than normal and was taken care of during procedure.No blood transfusion required.Resealed multiple times with ligasure to resolve the bleeding.The device was cleaned about 4-5 times after every time it came out of the patient.Also, knife blade could not retract and advance.Force had to be used to open and close the jaws.Jaws eventually opened after multiple trials.Opened another lxmj44s device as a remedy for the first device but had the reported issue as the first.An lf1944 device used to seal where the xp had failed, with the same generator with version 4.0.4.
|