• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Catalog Number 9-PFO-3025
Device Problem Difficult to Fold, Unfold or Collapse (1254)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2024
Event Type  malfunction  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6)2025, a 30-25mm amplatzer talisman pfo occluder was chosen for a patent foramen ovale (pfo) closure using a 9f amplatzer talisman delivery sheath.Heparin was administrated at the beginning of the procedure.During procedure, a thrombus was noted on the distal side of the delivery sheath.The thrombus was over 1cm and fiber-like shape.The sheath was in the patient for 2-3minutes.The physician was able to aspire the sheath but the right side of the device had a bulbous shape.The physician flushed the device after the aspiration of the thrombus but there was a small part of thrombus that couldn't be removed.A decision was taken to replace the device.The deformed device was never released from the delivery cable while inside the patient.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.A new 30mm amplatzer multi-fenestrated septal occluder - cribriform was chosen and completed the procedure.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure.The patient had been compliant with the prescribed anticoagulant.The patient was reported stable.
 
Manufacturer Narrative
An event of device deformity was reported.The investigation confirmed the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TALISMAN PFO OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18734253
MDR Text Key336286064
Report Number2135147-2024-00736
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9-PFO-3025
Device Lot Number8569324
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexMale
-
-