• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANDON MEDICAL CO.,LTD IHEALTH; Coronavirus antigen detection test system.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANDON MEDICAL CO.,LTD IHEALTH; Coronavirus antigen detection test system. Back to Search Results
Model Number ICO-3000
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2024
Event Type  malfunction  
Event Description
Event details: all of the tests in this package provided a false negative.It was ordered through amazon.I would like a return if possible.
 
Manufacturer Narrative
1)the customer used all 5 tests in the box and threw away all of the materials.The customer has not indicated why they think the tests results were false negative.Since these tests were ordered from amazon, an authorized agent of ihealth, there is no reason to believe that the tests were non-authentic.2)the customer has not responded to a request for additional information about why she thinks the results were false;the customer didn't pcr test ;results were compared using antigen reagents from different brands, but no relevant testing process or time was provided.3)test to lot: 231co20904 batch of product retention samples,the test result is qualified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IHEALTH
Type of Device
Coronavirus antigen detection test system.
Manufacturer (Section D)
ANDON MEDICAL CO.,LTD
tianjin, airport economic area 30030 8
CH  300308
Manufacturer (Section G)
ANDON MEDICAL CO.,LTD
tianjin, airport economic area 30030 8
CH   300308
Manufacturer Contact
ren jun
tianjin, airport economic area 30030-8
CH   300308
MDR Report Key18738202
MDR Text Key335823613
Report Number3008573045-2024-00003
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EUA210470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial
Report Date 02/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberICO-3000
Device Lot Number231CO20904
Was Device Available for Evaluation? No
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age30 YR
Patient SexFemale
-
-