The meter serial number is (b)(6).The product has not been received at this time and is not expected to be returned.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Section e3: occupation is patient/consumer.
|
Section b6 was updated.Per product labeling: coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine-based) origins are not a coaguchek-specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.The investigation did not identify a product problem.
|