C.R. BARD, INC. (BASD) -3006260740 SITE~RITE 8 ULTRASOUND SYSTEM CONSOLE WITH CONNECTOR; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Catalog Number 9770600 |
Device Problems
Poor Quality Image (1408); No Apparent Adverse Event (3189)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/26/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
H11: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and is pending evaluation.Results are expected soon.
|
|
Event Description
|
Sr8 is producing images with black line on top of image.Swapped probes.Same issue.No other information was provided.
|
|
Event Description
|
Sr8 is producing images with black line on top of image.Swapped probes.Same issue.No other information was provided.
|
|
Manufacturer Narrative
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was returned to the service facility for evaluation.During evaluation, the reported issue of sr8 is producing images with black line on top of image is unconfirmed.The sr8 was evaluated side by side with a test sr8 and the images are comparable.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.H3 other text : evaluation summary findings in h:11.
|
|